Compressed air requirements
For Leman Instruments nitrogen, zero air and oxygen generators.
Before installing a generator, check three things about the air that feeds it: its quality, its flow and its pressure. This page sums up what each Leman generator needs — and which ones need no compressed air at all.
Nitrogen, zero air and oxygen
Nitrogen, zero air and oxygen generators, and the DUO and TRIO LCMS stations, run on an external compressed-air supply.
Built-in compressor or water
Generators with a built-in compressor, and hydrogen generators (which run on deionised water), need no compressed air.
Section 1
Air quality
Oil-free, dried compressed air: ISO 8573-1:2010 Class 3 [3:3:3]
The standard sets limits for three contaminants — solid particles, water and oil. Leman Instruments generators require Class 3 for all three.
| Contaminant | Class 3 — required | Class 4 — for comparison |
|---|---|---|
| Solid particlesper m³ of air | 0.5 to 1 µm: max. 90 000 particles 1 to 5 µm: max. 1 000 particles | 1 to 5 µm: max. 10 000 particles |
| Waterpressure dew point | –20 °C or lower | +3 °C or lower |
| Oilliquid, aerosol and vapour | max. 1 mg/m³ | max. 5 mg/m³ |
Values at the reference conditions of the standard: 20 °C, 1 bar(a), 0 % relative humidity.
Fit static filters (particle, coalescing and activated carbon) and a dryer just before the generator's air inlet to reach Class 3. Air of a lower quality than Class 3 is not covered by the warranty.
Section 2
Air flow and pressure
The air flow needed is a multiple of the gas flow produced, and the air pressure must be higher than the outlet pressure of the generator.
| Generator | Air flow | Air pressure | Example |
|---|---|---|---|
| NITRO L · NITRO ELSD | × 5 | + 2 bar | 50 L/min air at 6 bar for 10 L/min N₂ at 4 bar |
| NITRO50 LS · NITRO ICP | × 7 | + 3 bar | 70 L/min air at 7 bar for 10 L/min N₂ at 4 bar |
| NITRO65 LS · NITRO60 LS | × 9 | + 3 bar | 90 L/min air at 7 bar for 10 L/min N₂ at 4 bar |
| NITRO20 · 30 · 40 LS2 (high flow) | up to × 4 | 6 to 10 bar at the inlet | 400 L/min air for 100 L/min N₂ |
| NITRO50 LS2 (high flow) | × 7 | + 4 bar | 700 L/min air at 8 bar for 100 L/min N₂ at 4 bar |
| NITRO65 LS2 (high flow) | × 9 | + 4 bar | 900 L/min air at 8 bar for 100 L/min N₂ at 4 bar |
| SOLO LCMS | × 4 | + 2 bar | 140 L/min air at 9 bar for 35 L/min N₂ at 7 bar |
| AERO50 (L, LS, LS2) | × 4 | + 1.5 bar | 12 L/min air at 5.5 bar for 3 L/min zero air at 4 bar |
| AERO FTIR · AERO TOC · AERO70 | × 5 | + 2 bar | 100 L/min air at 6 bar for 20 L/min zero air at 4 bar |
| OXY20 | × 30 | + 2 bar | 15 L/min air at 6 bar for 0.5 L/min O₂ at 4 bar |
| OXY50 | × 50 | + 3 bar | 25 L/min air at 7 bar for 0.5 L/min O₂ at 4 bar |
| DUO LCMS | × 3 | + 2 bar | 180 L/min air for 30 L/min N₂ + 30 L/min zero air |
| TRIO LCMS | × 3 to 7 | + 1 bar | 290 L/min air at 9 bar for all three gases at 8 bar |
For several generators on the same line, add up their air flows.
Section 3
Which compressor?
Any brand of compressor will do, or the laboratory air line, as long as the quality, flow and pressure above are respected.
For small bench-top generators
Not sure?
Send us your compressor models and the generators you plan: we check quality, flow and pressure for you.