Search
trio lcms sciex gas station
Nitrogen N2 molecule icon
Zero air molecule icon (N2 and O2)

TRIO LCMS-25/20/25

N2 + Zero Air Gas Station – Designed for the AB-SCIEX LC/MS Series

Leman Instruments designed the TRIO LCMS Gas Station to deliver, from a single unit, the Nitrogen and Air flows, purities and pressures required by the AB-SCIEX API 3200 to QTRAP 5500 LC/MS Series.

Compressed Air, supplied by an external source, is first filtered, then stored in a buffer tank.

A first single-stage Smart PSA unit extracts Nitrogen at >99.999% purity through a set of filters filled with carbon molecular sieve. A second single-stage Smart PSA unit dries and cleans the Air.

The three output lines — Source gas, Curtain gas and Exhaust gas — are each regulated to their set value and filtered of remaining particulates down to 0.1µm. All gases are free of phthalates and suspended liquid.

Each instrument is equipped with high performance communication interfaces RS485, USB and Ethernet to create a very flexible gas network with local or central control. Due to the software being focused on safety, automatic regulation and reliable communications, the TRIO LCMS is easy to install, reliable, safe and pleasant to operate.

Features

  • Reduces operation costs. Return on investment within 12 months maximum (exact estimate with our ROI calculator). Integrated Power Economy function.
  • Three gases from one unit: Curtain, Source and Exhaust gas for AB-SCIEX LC/MS.
  • Gases available 24/7 at constant purity. No phthalates, no liquid in suspension.
  • Very safe operation, internal leak-test, automatic shut-down, protection against over-temperature.
  • No handling, no storage, no cylinder rental fee.
  • No compressor, no consumables. Internal maintenance free.

Models

TRIO LCMS-25/20/25

Max outflow @1013 hPa / 20°C

  • TRIO LCMS-25/20/25 : Source gas: 25 NL/min Zero Air
  • TRIO LCMS-25/20/25 : Curtain gas: 20 NL/min N2
  • TRIO LCMS-25/20/25 : Exhaust gas: 25 NL/min dry air

Output purity

Source: dry Zero Air, <0.1ppm HC | Curtain: >99.999% N2, <0.1ppm HC, -45°C dew point | Exhaust: dry compressed air | Contaminants: phthalates and suspended liquid free | Particulates filtered down to 0.1µm on all three lines

Dew point

-45°C (Curtain N2)

Outlet pressure

Individually controlled and adjustable via touchscreen from 0.5 to 8 bar (7 to 116 psig)

Compressed air

Supplied by external compressed air supply

Inlet / outlet fittings

Output: BSP female 1/4″ cylindrical

Power supply

100 to 120VAC or 220 to 240VAC, 50 to 60 Hz

Power consumption

Maximum 500W

Manual control

Through a 7″ TFT-LCD color display with touch screen on the front panel. Display of major parameters, functioning status and alarms. Intuitive navigation to functions by menus and sub-menus.

Remote control

  • USB, RS485 (Mod-Bus) or Display
  • Through Ethernet 10/100 network
  • Logbook download by USB

Safety

Internal leak-test, automatic shut-down, protection against over-temperature

Sound pressure

<59 dB(A) @ 1m

Certification

CE, UL and CSA

Dimensions cm / in

  • L: 70 cm / W: 48 cm / H: 69 cm
  • L: 27.6in / W: 18.9in / H: 27.2in

Net weight

100 kg / 220 lbs

Format

Floor-standing casing