Search
nitro ls2 nitrogen generator
Nitrogen N2 molecule icon

NITRO40LS2

High Flow PSA Nitrogen Generator

Leman Instruments designed the NITRO40LS2 product line of Nitrogen Generators, to fit almost any type of application which needs a high flow of dry NITROGEN (> 99.99%) in a laboratory environment.

Dry and clean compressed air, supplied by an external source, is first filtered from particulates down to 5 µm.

Based on the field-proven Pressure Swing Adsorption (PSA) technology, nitrogen is then extracted from a set of filters filled with high-efficiency CMS. After the extracting process, N2 is filtered of remaining particulates down to 0.1 µm and pressure is regulated electronically to the set value.

O2 content can be monitored live. As an option, an O2 monitoring and alarm system can be installed.

Models with outflows of 50, 100, 200, 300, 500 and 750 NL/min N2 are available. Each instrument is equipped with modern and high-performance communication interfaces, RJ45 and USB as standard, to create a very flexible gas network with local or central control.

Features

  • Reduces operating costs – return on investment within 18 months.
  • N2 available 24/7 at constant purity. No contamination.
  • Independent source of dry N2 – needs only a dry oil-free compressed air supply and electricity; can easily be moved around the laboratory.
  • Low functioning noise (< 47 dBA).
  • Local remote control from PC; remote control via Internet (option).
  • Very safe operation: internal leak-test, automatic shut-down.
  • No handling, no storage, no cylinder rental fee.
  • Internally consumable free.

Models

  • NITRO40LS2-50
  • NITRO40LS2-100
  • NITRO40LS2-200
  • NITRO40LS2-300
  • NITRO40LS2-500
  • NITRO40LS2-750

Max outflow @1013 hPa / 20°C

  • NITRO40LS2-50 : 50 L/min
  • NITRO40LS2-100 : 100 L/min
  • NITRO40LS2-200 : 200 L/min
  • NITRO40LS2-300 : 300 L/min
  • NITRO40LS2-500 : 500 L/min
  • NITRO40LS2-750 : 750 L/min

Output purity

  • Minimum 99.99% N2
  • <-45°C dew point
  • particulates <0.1µm

Dew point

<-45°C

Outlet pressure

From 0.5 to 7 bar, adjustable by software (7 to 100 psig)

Air consumption

Maximum ratio of 4 – air flow consumption for N2 flow generated

Compressed air

Oil-free, dried compressed air; according to DIN ISO 8573-1 class 3

Inlet air pressure

6 bar (85 psig) minimum to 10 bar (140 psig) maximum

Inlet / outlet fittings

Inlet G1/2″ – outlet steel 3/8″ BNP

Power supply

Automatic switching from 90 to 260 VAC, 47 to 63 Hz

Power consumption

Maximum 150W

Manual control

Through a 7″ TFT-LCD colour display with touchpad on the front panel. Display of major parameters, functioning status and alarms. Intuitive navigation to functions by menus and sub-menus.

Remote control

  • RS485 or USB (Mod-Bus)
  • Through Ethernet 10/100 network

Safety

  • O2 content in the N2 outflow can be monitored live
  • optional O2 monitoring and alarm system

Sound pressure

< 47 dB(A)

Dimensions and weight

  • Dimensions: LS-50 & -100: W55 / H120 / D60 cm | LS-200 & -300: W55 / H220 / D80 cm | LS-500 & -750: W110 / H220 / D80 cm
  • Net weight (kg): LS-50: 75 | LS-100: 85 | LS-200: 142 | LS-300: 182 | LS-500: 325 | LS-750: 430

Format

Floor-standing cabinet