- Application Types
- Combustion gas for multi-detector GC/FID networks
- Dilution and zero gas for AQM calibrators
- Carrier gas for permeation tube systems
- Central air supply for multi-instrument laboratories
AERO50LS2
High Flow Ultra Zero Air Generator
Leman Instruments designed the AERO50LS2 product line of Ultra Zero Air Generators to fit almost any type of application which needs a high flow of Ultra High Purity Air in a laboratory environment.
Dry and clean compressed air, supplied by an external source, is first filtered from particulates down to 5µm and partially dried.
It then passes through a hot catalytic chamber to remove CO, CH4 and HC. Based on the field-proven Smart Pressure Swing Adsorption (SPSA) technology, air passes through a set of filters filled with molecular sieve and activated charcoal to be dried and cleaned from the last traces of HC.
After the purification process, air is filtered down to 0.01µm of remaining particulates and pressure is regulated electronically to the set value. Models with outflows of 50, 100, 200, 300, 500 and 750 NL/min air are available.
Each instrument is equipped with modern and high-performance communication interfaces, RJ45 and USB as standard, to create a very flexible gas network with local or central control.
Features
- Reduces operating costs – return on investment within 1.5 years.
- UHP Air available 24/7 at constant purity, improving resolution and detection limit.
- Independent source of air – needs only a dry oil-free compressed air supply and electricity.
- No handling, no storage, no cylinder rental fee, no consumables.
- Very safe operation: internal leak-test, automatic shut-down, protection against over-temperature.
- Local and remote control via Ethernet RJ45, standard on every unit.
Gas
Models |
|
Max outflow @1013 hPa / 20°C |
|
Output purity |
|
Dew point | <10 ppm H2O |
Outlet pressure | From 0.5 to 7 bar, adjustable by software (7 to 100 psig) |
Air consumption | Maximum ratio of 4 – air flow consumption for air flow generated |
Site & Installation
Compressed air | External supply – oil-free dried compressed air according to DIN ISO 8573-1 class 3 |
Inlet air pressure | At least 1 bar over the set outlet pressure to Maximum 14 bar |
Inlet / outlet fittings | Inlet G1/2″ – outlet stainless steel 3/8″ BNP |
Power supply | Automatic switching from 90 to 260 VAC, 47 to 63 Hz |
Power consumption | Maximum 150W |
Manual control | Through a 7″ TFT-LCD colour display with touchpad on the front panel. Display of major parameters, functioning status and alarms. Intuitive navigation to functions by menus and sub-menus. |
Remote control |
|
Dimensions & Weight
Dimensions and weight |
|
Format | Floor-standing cabinet |