Search
aero ls2 zero air generator
Zero air molecule icon (N2 and O2)

AERO TOC LS2

High Flow UHP Air Generator for TOC/THC Analyzers

Leman Instruments designed the AERO TOC LS2 product line of UHP Air Generators for TOC/THC analyzers to fit almost any type of application which needs a high flow of Ultra High Purity Air in a laboratory environment.

Compressed air, supplied by an external source, is first filtered from CO, CH4 and HC; then, based on the field-proven Pressure Swing Adsorption (PSA) technology, CO2, moisture and remaining HC are removed.

After the purification process, UHP air is filtered of remaining particulates and pressure is regulated electronically to the set value. Models with outflows of 50, 100, 200, 300, 500 and 750 NL/min air are available.

Each instrument is equipped with modern and high-performance communication interfaces, RJ45 and USB as standard, to create a very flexible gas network with local or central control.

Features

  • Reduces operating costs – return on investment within 1 year.
  • UHP Air available 24/7 at constant purity, improving resolution and detection limit.
  • Independent source of air – needs only a dry oil-free compressed air supply and electricity.
  • No handling, no storage, no cylinder rental fee, no maintenance or consumables.
  • Very safe operation: internal leak-test, automatic shut-down, protection against over-temperature.
  • Local and remote control via Ethernet RJ45, standard on every unit.

Models

  • AERO TOC LS2-50
  • AERO TOC LS2-100
  • AERO TOC LS2-200
  • AERO TOC LS2-300
  • AERO TOC LS2-500
  • AERO TOC LS2-750

Max outflow @1013 hPa / 20°C

  • AERO TOC LS2-50 : 50 L/min
  • AERO TOC LS2-100 : 100 L/min
  • AERO TOC LS2-200 : 200 L/min
  • AERO TOC LS2-300 : 300 L/min
  • AERO TOC LS2-500 : 500 L/min
  • AERO TOC LS2-750 : 750 L/min

Output purity

  • <5ppm H2O
  • <1ppm CO2
  • <0.25ppm HC
  • <0.25ppm CO and CH4
  • particulates <0.01µm

Dew point

-60 °C

Outlet pressure

From 0.5 to 7 bar, adjustable by software (7 to 100 psig)

Air consumption

Maximum ratio of 5 – air flow consumption for air flow generated

Compressed air

External supply – oil-free dried compressed air according to DIN ISO 8573-1 class 3

Inlet air pressure

At least 2 bar over the set outlet pressure to Maximum 14 bar

Inlet / outlet fittings

Inlet G1/2″ – outlet stainless steel 3/8″ BNP

Power supply

Automatic switching from 90 to 260 VAC, 47 to 63 Hz

Power consumption

Maximum 150W

Manual control

Through a 7″ TFT-LCD colour display with touchpad on the front panel. Display of major parameters, functioning status and alarms. Intuitive navigation to functions by menus and sub-menus.

Remote control

  • RS485 or USB (Mod-Bus)
  • Through Ethernet 10/100 network

Dimensions and weight

  • Dimensions: LS-50 & -100: W55 / H120 / D60 cm | LS-200 & -300: W55 / H220 / D80 cm | LS-500 & -750: W110 / H220 / D80 cm
  • Net weight (kg): LS-50: 75 | LS-100: 85 | LS-200: 142 | LS-300: 182 | LS-500: 325 | LS-750: 430

Format

Floor-standing cabinet