DUO LCMS-35/50
Nitrogen & Zero Air Generator – Designed for the Bruker EVOQ Triple Quad LC-MS/MS
LEMAN Instruments designed this DUO LCMS variant specifically to match the gas demand of the Bruker EVOQ Triple Quadrupole LC-MS/MS series, delivering nitrogen and zero air simultaneously at the flow and pressure the instrument actually draws.
Dry and clean compressed Air, supplied by an external source is first filtered from particulates down to 5µm.
Based on the field proven Pressure Swing Adsorption (PSA) technology, Nitrogen is then extracted from a set of filters filled up with high efficiency CMS. After the extracting process, N2 is filtered of remaining particulates down to 0.01 µm and pressure is regulated electronically to the set value, between 0.5 and 8 bar (7 to 116 psig).
Same process for the Zero Air. Model with outflow of up to 35 NL/min N2 and up to 50 NL/min ZA simultaneously is available, matched to the Bruker EVOQ gas requirement of 32 L/min N2 and 50 L/min dry air, with headroom on the nitrogen side.
Each instrument is equipped with modern and high-performance communication interfaces Ethernet as standard to create a very flexible gas network with local or central control. Due to the software being focused on safety, automatic regulation, and reliable communications, the DUO LCMS-35/50 is easy to install, reliable, safe and pleasant to operate.
Features
- Reduces operation costs. Return on investment within 12 months maximum (exact estimate with our ROI calculator).
- N2 and ZA available 24/7 at constant purity. No contamination.
- Independent source of dry N2 and ZA. Needs only dry oil-free compressed air supply and electricity. Can be easily moved around the laboratory.
- Low functioning noise.
- Local remote-control via Ethernet.
- Very safe operation, internal leak-test, automatic shut-down.
- No handling, no storage, no cylinder rental fee.
- No compressor, no consumables. Internal maintenance free.
Gas
Models | DUO LCMS-35/50 |
Max outflow @1013 hPa / 20°C |
|
Output purity | Output N2 purity: 99% N2, -45°C dew point, particulates <0.01µm | Output ZA purity: -50°C dew point, CO2, THC, CH4 max. 10ppm each |
Dew point | -45°C (N2) / -50°C (Zero Air) |
Outlet pressure | 0.5 to 8 bar, adjustable directly on touchscreen for both gases independently (7 to 116 psig) |
Air consumption | Maximum ratio of 3 – air flow consumption / N2 flow generated; same for ZA |
Site & Installation
Compressed air | Oil-free, dried compressed air; DIN ISO 8573-1 class 3 |
Inlet / outlet fittings | Inlet G3/8″; outlet stainless steel 1/4″ compression (other on request) |
Power supply | Automatic switching from 90 to 260 VAC, 47 to 63 Hz |
Power consumption | Max 450W |
Manual control | Through a 7″ TFT-LCD color display with touchscreen, located on the front panel. Display of major parameters, functioning status and alarms. Intuitive navigation to functions by menus and sub-menus. |
Remote control | Through Ethernet 10/100 network |
Safety | Internal leak-test, automatic shut-down |
Certification | CE, UL and CSA |
Dimensions & Weight
Dimensions cm / in |
|
Net weight | 95 kg / 209 lbs |
Format | Floor-standing casing |