Search
DUO LCMS nitrogen and zero air generator with Air and N2 touchscreen
Nitrogen N2 molecule icon
Zero air molecule icon (N2 and O2)

DUO LCMS-20/70

Nitrogen & Zero Air Generator – Designed for the PerkinElmer QSight LC-MS/MS (Single or Dual Source)

LEMAN Instruments designed this DUO LCMS variant specifically to match the gas demand of the PerkinElmer QSight Triple Quad LC-MS/MS series (single or dual source), delivering nitrogen and zero air simultaneously at the flow and pressure the instrument actually draws.

Dry and clean compressed Air, supplied by an external source is first filtered from particulates down to 5µm.

Based on the field proven Pressure Swing Adsorption (PSA) technology, Nitrogen is then extracted from a set of filters filled up with high efficiency CMS. After the extracting process, N2 is filtered of remaining particulates down to 0.01 µm and pressure is regulated electronically to the set value, between 0.5 and 8 bar (7 to 116 psig).

Same process for the Zero Air. Model with outflow of up to 20 NL/min N2 and up to 70 NL/min ZA simultaneously is available, matched exactly to the PerkinElmer QSight gas requirement of 20 L/min N2 and 70 L/min dry air.

Each instrument is equipped with modern and high-performance communication interfaces Ethernet as standard to create a very flexible gas network with local or central control. Due to the software being focused on safety, automatic regulation, and reliable communications, the DUO LCMS-20/70 is easy to install, reliable, safe and pleasant to operate.

Features

  • Reduces operation costs. Return on investment within 12 months maximum (exact estimate with our ROI calculator).
  • N2 and ZA available 24/7 at constant purity. No contamination.
  • Independent source of dry N2 and ZA. Needs only dry oil-free compressed air supply and electricity. Can be easily moved around the laboratory.
  • Low functioning noise.
  • Local remote-control via Ethernet.
  • Very safe operation, internal leak-test, automatic shut-down.
  • No handling, no storage, no cylinder rental fee.
  • No compressor, no consumables. Internal maintenance free.

Models

DUO LCMS-20/70

Max outflow @1013 hPa / 20°C

  • DUO LCMS-20/70 : up to 20 NL/min
  • DUO LCMS-20/70 : up to 70 NL/min

Output purity

Output N2 purity: 99% N2, -45°C dew point, particulates <0.01µm | Output ZA purity: -50°C dew point, CO2, THC, CH4 max. 10ppm each

Dew point

-45°C (N2) / -50°C (Zero Air)

Outlet pressure

0.5 to 8 bar, adjustable directly on touchscreen for both gases independently (7 to 116 psig)

Air consumption

Maximum ratio of 3 – air flow consumption / N2 flow generated; same for ZA

Compressed air

Oil-free, dried compressed air; DIN ISO 8573-1 class 3

Inlet / outlet fittings

Inlet G3/8″; outlet stainless steel 1/4″ compression (other on request)

Power supply

Automatic switching from 90 to 260 VAC, 47 to 63 Hz

Power consumption

Max 500W

Manual control

Through a 7″ TFT-LCD color display with touchscreen, located on the front panel. Display of major parameters, functioning status and alarms. Intuitive navigation to functions by menus and sub-menus.

Remote control

Through Ethernet 10/100 network

Safety

Internal leak-test, automatic shut-down

Certification

CE, UL and CSA

Dimensions cm / in

  • L: 70 cm / W: 48 cm / H: 69 cm
  • L: 27.6in / W: 18.9in / H: 27.2in

Net weight

100 kg / 220 lbs

Format

Floor-standing casing