The Leman generator for your LC-MS
Nitrogen and zero air for LC-MS systems from 11 makers — one generator per instrument, or one central system for the whole lab.
TRIO LCMSBrowse by instrument family
Triple quads, Q-TOFs, Orbitraps and the rest: each family has its own nitrogen profile.
Browse by maker
SCIEX, Agilent, Waters, Thermo Fisher, Shimadzu, Bruker, PerkinElmer and more, current systems and older models still widely installed.
All instruments
| Maker | Instrument | Gases | Leman generator | Status |
|---|
Sizing in one line
- An LC-MS source uses 10 to 45 L/min of nitrogen, mostly at 99 % purity.
- Some systems also need zero or dry air: SCIEX, Shimadzu LCMS-80x0, Bruker EVOQ, PerkinElmer QSight.
- Collision gas is a few mL/min: 99.999 % nitrogen on Agilent and Thermo Orbitrap, argon on Waters, Shimadzu and Thermo TSQ.
- For several instruments, one NITRO LS2 feeds the whole lab from 50 to 750 L/min.
Built for the big names
The TRIO LCMS supplies the three SCIEX gases from one station. The DUO LCMS comes in versions matched to Bruker EVOQ and PerkinElmer QSight.
No cylinders, no dewars, and nitrogen at the right pressure 24/7.
Instrument names and trademarks belong to their owners. This list reflects Leman Instruments' assessment of each instrument's gas requirements, based on the makers' published documentation and our installations. It does not imply approval by the instrument makers. "Ask us": the maker does not publish gas figures for that model; we confirm them with you at quotation. Helium and argon, where an instrument needs them, are not produced by Leman generators.